memory locations are indexed, the display will indicate 00.0. Subsequent
operation of a number key and M returns to the first L

eq

result again.

Memory capacity is approximately 720 periods (reduced to approx. 499
in B version).

Environmental Mode

Answers are selected in this mode by using the same key codes as for
the F key. Therefore 1M brings the first L

eq

result from the memory to

the display, with subsequent operation of the M key stepping through the
L

eq

results in time order until all the L

eq

results have been displayed.

Subsequent operation of the sequence 1M returns to the first L

eq

result.

Similarly, the sequence 4M selects the L

10

result for the first period,

which may then be stepped through by subsequent operation of the M
key.

Memory capacity is approximately 90 periods (reduced to approx. 60 in
B version).

Event Mode

Operation of any numerical key followed by M will bring the time in hours
and minutes of the oldest event in the memory to the display. Subsequent
operations of the M key display the SEL (L

AX

) of the event, then its

maximum value, and finally its duration in seconds.

Continued operation of the M key displays the time of the second oldest
event, followed by its SEL, maximum, and duration, and so on. Re-op-
eration of a numeral and M returns to the oldest event again.

Memory capacity is approximately 180 events (reduced to approx. 119
in B version).

Analysis Mode

/24.

393 Handbook