15.

After the delayed start has been reached, the first L

10

index

(calculated for the period between 6.00 and 7.00 am) may be
obtained by pressing 4M (sequence for L

10

).

16.

Press M to obtain the second L

10

sample.

17.

Press M again to obtain subsequent samples.

If the instrument is collected on the day after the measurements were
taken, only the first 18 answers from the L

10

memory will be required.

The calculation of the full L

10

18 hour index is achieved by adding them

together arithmetically and then dividing the sum by 18.

If necessary, the period L

eq

, L

max

, L

50

, and L

90

answers may also be

accessed by keying in the relevant codes (1M, 0M, 5M, and 6M respec-
tively).

All results may be hand written on pre-printed report sheets (SP0884),
or if the CEL-393B with built in interface is used, they can be printed
automatically using the CEL-238, CEL-338 and CEL-438 Digital Proces-
sor/Printer.

Switch the instrument off when all relevant answers have been obtained.

6.3

Event Mode

This mode is used for measuring peak air over-pressures that occur at
comparatively random intervals in sensitive locations, where possible
claims for noise nuisance may occur.

Requirement:

Monitor peak levels around a site where quarrying operations and
blasting take place.

Proceed as follows.

/66.