Casella CEL CEL-393 User Manual

Page 92

Advertising
background image

Pre and post overload detectors with memory.

Calibration

Field calibration is achieved by means of the CEL-284/2
Precision Acoustic Calibrator.

Temperature Range

-10 to +50

o

C.

Dimensions

285 x 75 x 25mm.

Weight (less battery)

800g, 1.75lb, (A version).

Measurement Ranges

1

Note

1

These figures were measured for A and B versions. For C and D versions, the
90dB range is not available, while all other ranges will be similar.

Note

2

Lower reading is quoted at +10dB on noise floor. Upper reading allows crest
factor 10.

Reference Information Required by IEC 651 and 804

(section numbers relate to IEC 651 section 11.2)

1.

A 35mV/Pa condenser microphone type CEL-186 is required
(or 80mV/Pa type CEL-190 for C and D models) and should
be screwed no more than finger-tight to the end of the pream-
plifier.

2.

The reference direction of incidence is perpendicular to the
microphone diaphragm.

/88.

393 Handbook