Casella CEL CEL-393 User Manual

Page 94

Advertising
background image

14)

The recommended windscreen has negligible effect at frequen-
cies <12.5kHz.

15.

The instrument should be calibrated with either a CEL-284/2
electroacoustic calibrator or an RFT 05001 pistonphone.

16.

For maximum accuracy the operator and instrument case
should be more than 1m behind the microphone.

17.

The CEL-393 is designed to be used only with its own internal
filter set. Models containing filters are designated CEL-393B
and CEL-393D.

18.

The electrical impedance of equipment attached to the log DC
or conditioned AC output connectors should be >3k3 Ohms.
Digital outputs are available for CEL-238 Secondary Proces-
sor or CEL-433 Serial Interface.

19.

The reference frequency is 1kHz.

20.

The reference range is '70' ('60' on C and D versions).

/90.

393 Handbook