The following outputs are provided in all versions of the instrument.

AC

Warning

Do not insert two-pole plugs into this socket as they
could damage components located immediately be-
hind it.

This is a conditioned AC output from an impedance of 3k3 ohms at a
level of 2.3V RMS for full scale deflection. It is a standard single pole
2.5mm jack socket located on the side of the instrument (see Figure 3).

DC

A log DC output with a nominal value of 75mV/dB is provided on pin 10
of the connector (see Figure 7). This output also has a source impedance
of 3k3 ohms.

Figure 7 Auxiliary socket, pin number & function

890019

1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.

Out 3
Stop paper
Stop pen
Not connected
Not connected
0V (ground)
Ext. unit 1
Ext. unit 2
Out 4
DC output
Not connected
+9V input
Not connected
Step pulse
Pause

8

7

6

5

4

3

2

1

13

12 11 10

9

14

15

/36.

393 Handbook