Erial number and calibration date, Serial number and calibration date – Xantrex Technology XMP 2600 User Manual

Page 108

Advertising
background image

Power Supply setup

Power module setup

Serial number and calibration date

Select the power module (move the cursor to the condensed status symbol of the
power module) and press the RCL key. The illustration, below, shows a typical
display.

Last value of
event status byte

system

symbol

system

parameters to set

condensed

modules status

Module’s

serial number

Calibration

date

Installed

options


Setup mode display – module’s controller information

Information shown

Description

Serial number

Power module serial number.

Calibration date

Last power module calibration date: dd/mm/yy.

Installed options

A HEX number representing the power module options
byte:

Bit 4.... cleared when the Power Module has output

disconnect relay.

Bit 5.... cleared if the Power Module has polarity

reversal relays.

Bit 3.... set if this module has power monitoring

logic (e.g. 1/8

wide

36V/40A module)

Event status byte

A HEX number representing the event register of the power
module’s status structure:





Bit 7: PON : set when a Power On event occurs.
Bit 5: CMD : set when the Power Module detected

an error while communicating with the XMP
2600 Main Controller..

Bit 4: OUT : the Output Register had a bit changed.
Bit 3: ERR : set when an Error event occurred.
Bit 2: FLT : the Faults Register had a bit changed.
Bit 1: WRN : the Warnings Register had a bit

changed.

Bit 0: OPC : Operation Complete - the Power

Module finished processing the XMP 2600
Main Controller Command. For commands
that effect the output of the Power Module,
the OPC flag is set when the output settled to
within 2% of full scale from the programmed
value.

To clear the display, press the SYSTEM key or the LEFT / RIGHT selection keys.

To view more information (shown below) press the RCL key a second time.

XMP 2600 User’s Guide

rev. 1.0

89