Isplayed entry guidance, Limit value, Imit indication – Xantrex Technology XMP 2600 User Manual

Page 65: Displayed entry guidance (limit value), Limit indication

Advertising
background image

Programming power module parameters

Basic operation

Parameter

Value

Lower limit

Upper limit

Voltage Vset

Vmax

LIM

Current Iset

Imax

Voltage Vset

Vmax

HIGH

Current Iset

Imax

Voltage 0V

Vset

LOW

Current 0A

Iset

Displayed entry guidance (limit value)

When you program a value for a parameter that has voltage and current values, the
right side of the display’s upper row shows the most relevant limit on the allowed
value to program. The table below summarizes the displayed limiting values:

Parameter

Value

Displayed limit

Remarks

Voltage

Upper: LM or
OV

The lower of Vlim or OVset

SET

Current

Upper: LM or
OC

The lower of Ilim or OCset

Voltage Lower:

SE

Vset

PROT

Current Lower:

SE

Iset

Voltage Upper:

MX Vmax

LIM

Current Upper:

MX Imax

Voltage Lower:

SE

Vset

HIGH

Current Lower:

SE

Iset

Voltage Upper:

SE

Vset

LOW

Current Upper:

SE

Iset

Limit indication

When you try to program a value outside its allowed range, the value is not accepted
and a warning message (“*** VALUE LIMIT ***”) is displayed on the display’s
lower row.

If the internal buzzer is set to “always” (see the chapter on power supply setup) – a
short beep will be heard as well.

46

XMP 2600 User’s Guide

rev. 1.0