Arameters value limits, Ower and upper limits, Parameters value limits – Xantrex Technology XMP 2600 User Manual

Page 64: Lower and upper limits

Advertising
background image

Basic operation

Programming power module parameters

Parameter

Values to program (left to right)

SET

Output voltage and current limit (or voltage limit and

output current if operating as a current source).

Current limiting scheme (Foldback selection).

RIPL

Simulated ripple at the module’s output: level (OFF, LOW

or HIGH) and frequency (50Hz, 60Hz, 100Hz,
120Hz, 400Hz, 800Hz, 2KHz).

PROT


Over voltage and over current protection threshold levels.
Protection levels programming method: AUT=automatic

(auto-tracking) or MAN=manual (user
programmed).

MAX

(none – shown for reference only)

LIM

Voltage and current values that serve as limits on the

allowed output values to program (SET parameter).

HIGH

Voltage and current thresholds for the high voltage and

current portion of the window warning mechanism.

State of the high voltage and current window warning

mechanism: ON or OFF.

LOW

Voltage and current thresholds for the low voltage and

current portion of the window warning mechanism.

State of the low voltage and current window warning

mechanism: ON or OFF.

SYNC

Selection of a module condition that will cause a SYNC

signal to be generated. The SYNC signals from all
the modules are summed to produce the output at
the rear of the XMP’s mainframe.

TRIG

Selection of the module’s response to a hardware or

software trigger. Hardware trigger signal can be fed
at the rear of the XMP’s mainframe. The signal is
distributed to all the power modules in the
mainframe.

Parameters

value limits

Voltage and current values associated with parameters (e.g. SET, PROT, HIGH)
have upper and lower limits on their allowed value.

Lower and upper limits

The following table lists the upper and lower limiting values for each parameter:

Parameter

Value

Lower limit

Upper limit

Voltage

0V

The lower of: Vmax, OVset, Vlim

SET

Current

Imin

The lower of: Imax, OCset, Ilim

Voltage

Vset

Vmax + 10%

PROT

Current Iset

Imax

+10%

XMP 2600 User’s Guide