Installation

Location, Mounting And Ventilation

Use the XMP 2600 power supply in rack-mounted applications only. The XMP
2600 is designed to fit into standard 19” racks.

Install the XMP 2600 using only rack-mounted telescopic sliding rails, or rack-
mounted slide brackets. Note that the front panel is not designed to hold the unit’s
weight.

Fasten the unit to the rack using bolts on both sides of the unit’s front panel.

Keep an unrestricted air space of at least 5cm (2”) at the rear so the exhausted hot air
can flow freely.

Rear Panel Description

The rear panel contains all the connectors available to the user. It, also, serves as the
ventilation exhaust of the unit.

IEEE-488

RS-232

NET 2

NET 1

TRIG IN

SYNC OUT

ON-OFF CTRL

S8

S7

S6

S5

S4

S3

S2

S1

Module’s output connectors

Net

expansion

link

GPIB address

and default

control link

dip-switch

GPIB

RS232

Trigger

input

Sync

output

ON / OFF

control

. . . . .

GPIB

Address and

default

control link

The GPIB (IEEE-488) address and the default control link can be changed from the
rear panel using the switches labeled "IEEE-488 ADDR". They can also be changed
using the front panel but that change will be temporary.

Switch number

function

8 (left most)

Should always be UP.

7

Enable the GPIB link for remote control when UP.

6

Enable the RS232 link when UP.

If the GPIB link is enabled then the RS232 link will be in
Monitor mode at Power On. If the GPIB link is disabled then the
RS232 link will be in Control mode at Power On.

5 to 1

GPIB address (see table below).

Changes in switch position are recognized when the XMP 2600 is
turned on or undergoes full reset.

6

XMP 2600 User’s Guide

rev. 1.0