What to do if...

ProLab 2000

100

ba75577e05

11/2013

No stable measured

value

Obviously incorrect

measured values

Buffer solutions:

– Incorrect buffer solutions

– Change buffer solutions

– Buffer solutions too old

– Use only once.

Note the shelf life

– Buffer solutions depleted

– Change buffer solutions

Cause

Remedy

Cause

Remedy

pH electrode:

– Junction contaminated

– Clean junction

– Membrane contaminated

– Clean membrane

Test sample:

– pH value not stable

– Measure with air excluded if

necessary

– Temperature not stable

– Adjust temperature if

necessary

Electrode + test sample:

– Conductivity too low

– Use a suitable electrode

– Temperature too high

– Use a suitable electrode

– Organic liquids

– Use a suitable electrode

Cause

Remedy

pH electrode:

– Not connected

– Connect electrode

– Cable broken

– Replace electrode

– pH electrode unsuitable

– Use a suitable electrode

– Temperature difference

between buffer and test
sample too high

– Adjust temperature of buffer

or sample solutions