Xylem ProLab 2000 User Manual
Page 100

What to do if...
ProLab 2000
100
ba75577e05
11/2013
No stable measured
value
Obviously incorrect
measured values
Buffer solutions:
– Incorrect buffer solutions
–
– Change buffer solutions
– Buffer solutions too old
– Use only once.
Note the shelf life
– Buffer solutions depleted
– Change buffer solutions
Cause
Remedy
Cause
Remedy
pH electrode:
– Junction contaminated
– Clean junction
– Membrane contaminated
– Clean membrane
Test sample:
– pH value not stable
– Measure with air excluded if
necessary
– Temperature not stable
– Adjust temperature if
necessary
Electrode + test sample:
– Conductivity too low
– Use a suitable electrode
– Temperature too high
– Use a suitable electrode
– Organic liquids
– Use a suitable electrode
Cause
Remedy
pH electrode:
– Not connected
– Connect electrode
– Cable broken
– Replace electrode
– pH electrode unsuitable
– Use a suitable electrode
– Temperature difference
between buffer and test
sample too high
– Adjust temperature of buffer
or sample solutions