Operation
ProLab 2000
52
ba75577e05
11/2013
Note
After a five-point calibration the calibration procedure is automatically
terminated. The menu item Terminate with ... is not displayed.
Note
The calibration line is calculated by linear regression.
18
Switch to measurement of the next buffer with <MENU/OK>.
or
When all buffers have been measured, use <
▲><▼> to select
the Terminate with ... menu item and confirm with <MENU/
OK>.
The calibration procedure is terminated.
The new calibration values are displayed as a message ([i]).
You have the following options:
Accept the new calibration values with <MENU/OK>.
Subsequently, the calibration record is displayed and
downloaded to the interface at the same time.
To switch to the measured value display without accepting
the new calibration values, press <MODE> or <ESC>.
[i] Buffer value = 4.010
[i] U = 184.0 mV
[i] Temperature =
24.8 °C
Terminate with ...
Terminate stability control
pH <CAL>