Calibration via a

Comparison meas.

For this calibration procedure, the Comparison meas. setting must be
set to On in the measuring menu.

Note
Before calibrating via a comparison measurement, the sensor should
be calibrated in the air calibration vessel.

Proceed as follows to calibrate the meter:

8

You have the following options:
 Accept the new calibration values with <MENU/OK>.

Subsequently, the calibration record is displayed and
downloaded to the interface at the same time.

 To switch to the measured value display without accepting

the new calibration values, press <MODE> or <ESC>.

[i] Relative slope = 0.98
i Sensor+++

Accept

LFOx1400-ID <CAL>

1

Connect the D.O. sensor to the meter.

2

Immerse the D.O. sensor in the reference solution.

3

In the measured value display, select the D.O. measuring
window with <

▲><▼> and <MODE>.

4

Start the calibration with <CAL>.
The calibration display appears.

[i] Immerse sensor in ref. sol.

Continue

LFOx1400-ID <CAL>