Xylem ProLab 2000 User Manual

Page 67

Advertising
background image

ProLab 2000

Operation

67

ba75577e05

11/2013

Temp. comp. (TC) /
Linear coeff.

0.000 ...
3.000 %/K

Coefficient of the linear
temperature compensation.

This setting is only available
when the linear temperature
compensation is set.

TDS factor

0.40 ... 1.00

TDS factor

This setting is only available
when the TDS display is set.

Measuring cell

Measuring cell used

Cal

Measuring cells the cell
constant of which is
determined by calibration in
the KCL control standard
solution (Calibration ranges
see page 68 and page 108).
The currently valid cell
constant is displayed in the
status line.

C0.01

Measuring cell with a nominal
cell constant of 0.010 cm

-1

.

The cell constant is
permanently set.

LF313T

Measuring cell LF313T(-ID),
nominal cell constant
0.100 cm

-1

.

The cell constant can be
adjusted in the range 0.090 to
0.110 cm

-1

.

man

Any measuring cells with
freely adjustable cell
constants in the range 0.250
to 25.000 cm

-1

.

Cell constant

0.090 to
0.110 cm

-1

Display and setting option for
the cell constant of the
measuring cell LF313T(-ID).

Man. cell const.

0.250 to
25.000 cm

-1

Display and setting option of
the cell constant of any
measuring cells (man).

Menu item

Possible
setting

Description