Xylem ProLab 2000 User Manual
Page 67

ProLab 2000
Operation
67
ba75577e05
11/2013
Temp. comp. (TC) /
Linear coeff.
0.000 ...
3.000 %/K
Coefficient of the linear
temperature compensation.
This setting is only available
when the linear temperature
compensation is set.
TDS factor
0.40 ... 1.00
TDS factor
This setting is only available
when the TDS display is set.
Measuring cell
Measuring cell used
Cal
Measuring cells the cell
constant of which is
determined by calibration in
the KCL control standard
solution (Calibration ranges
see page 68 and page 108).
The currently valid cell
constant is displayed in the
status line.
C0.01
Measuring cell with a nominal
cell constant of 0.010 cm
-1
.
The cell constant is
permanently set.
LF313T
Measuring cell LF313T(-ID),
nominal cell constant
0.100 cm
-1
.
The cell constant can be
adjusted in the range 0.090 to
0.110 cm
-1
.
man
Any measuring cells with
freely adjustable cell
constants in the range 0.250
to 25.000 cm
-1
.
Cell constant
0.090 to
0.110 cm
-1
Display and setting option for
the cell constant of the
measuring cell LF313T(-ID).
Man. cell const.
0.250 to
25.000 cm
-1
Display and setting option of
the cell constant of any
measuring cells (man).
Menu item
Possible
setting
Description