Xylem ProLab 2000 User Manual

Page 73

Advertising
background image

ProLab 2000

Operation

73

ba75577e05

11/2013

Note
You can terminate prematurely the manual Stability control function
with <MENU/OK> manually at any time. If the manual Stability control
function is terminated prematurely, the current measurement data is
not downloaded to the interface.

Criteria

With identical measurement conditions, the following applies:

2

Freeze the measured value with <SC>.
The display switches to the display of the selected channel as
necessary.
In the status line, [SC] is displayed.

3

With <MENU/OK> activate the Stability control function.
[SC] flashes while the stability control is active.
As soon as a stable measured value is recognized, the current
measurement data is downloaded to the interface. The
measured values of all channels meeting the criterion for
stability control are marked by SC.

4

Using <SC> or <MODE>, release the frozen measured value.
The [SC] status display disappears. The display switches to the
previous indication.

Measuring mode

Reproducibility

Response time

D.O. concentration

Better than 0.05 mg/l

> 10 seconds

D.O. saturation

Better than 0.6 %

> 10 seconds

D.O. partial
pressure

Better than 1.2 mbar

> 10 seconds