Xylem ProLab 2000 User Manual
Page 79

ProLab 2000
Operation
79
ba75577e05
11/2013
5
Press <MENU/OK>.
The measurement with stability control to determine the D.O.
concentration starts.
6
Wait for the end of the measurement with Stability control or
accept the measured value with <MENU/OK>.
The determined D.O. content is displayed as a message (
i
).
7
Open the setting of the correction factor with <MENU/OK>.
8
Using <
▲><▼>, set the correction factor to adjust the
displayed concentration value to the nominal value (value of
the comparison measurement). Subsequently, accept the
correction factor with <MENU/OK>.
9
Select Accept with <
▲><▼> and press <MENU/OK>.
The meter switches to the measured value display. The
[Factor] indicator appears in the measuring window.
[i] Concentration = 8.5 mg/l
[i] Temperature =24.8 °C
Terminate stability control
LFOx1400-ID <CAL>
[i] Concentration = 8.49 mg/l
[i] Temperature =24.8 °C
Set factor:
1.000
Accept
LFOx1400-ID <CAL>