2 measuring the ph value, Measuring the ph value – Xylem ProLab 2000 User Manual

Page 37

Advertising
background image

ProLab 2000

Operation

37

ba75577e05

11/2013

4.5.2

Measuring the pH value

Stability control

The Stability control function checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of
measured values. The display of the measured parameter flashes until
a stable measured value is available.

Independent of the setting for automatic Stability control (see page 35)
in the System menu, you can start the Stability control function
manually at any time.

Note
You can terminate prematurely the Stability control function with
<MENU/OK> manually at any time. If the Stability control function is
terminated prematurely, the current measurement data is not
downloaded to the interface.

1

Perform the preparatory activities (see page 36).

2

Immerse the pH electrode in the test sample.

3

Select the pH or mV display with <MODE>.

pH

6.949

24.8

°C

26.04.06 10:37

1

If necessary, select a channel (measured value) with <

▲><▼>.

2

Freeze the measured value with <SC>.
The display switches to the display of the selected channel as
necessary.
In the status line, [SC] is displayed.

3

With <MENU/OK> activate the Stability control function.
[SC] flashes while the stability control is active.
As soon as a stable measured value is recognized, the current
measurement data is downloaded to the interface. The
measured values of all channels meeting the criterion for
stability control are marked by SC.