4 dissolved oxygen, Dissolved oxygen – Xylem ProLab 2000 User Manual

Page 109

Advertising
background image

ProLab 2000

Technical data

109

ba75577e05

11/2013

Stand March 3, 2010

Translation of the legally binding German version

Note
The measuring ranges and accuracy values specified here apply
exclusively to the meter. The accuracy of the sensors has also to be
taken into account.

7.2.4

Dissolved oxygen

Accuracy (± 1 digit)

Variable

Accuracy

Temperature of the test
sample

/ Temperature compensation

None (Off)

± 0.5 %

Nonlinear (nLF)

± 0.5 %

0 °C ... + 35 °C
according to EN 27 888

± 0.5 %

+ 35 °C ... + 50 °C
Extended nLF function
according to WTW
measurements

Linear (lin)

± 0.5 %

+ 10 °C ... + 75 °C

SAL / range

0.0 ... 42.0

± 0.1

+ 5 °C ... + 25 °C

± 0.2

+ 25 °C ... + 30 °C

TDS [mg/l]

± 1

T [°C] / temperature sensor

NTC 30

± 0.2

0 °C ... + 55 °C

PT 1000

± 0.3

0 °C ... + 55 °C

Measuring ranges,

resolution

Variable

Measuring range

Resolution

D.O. concentration
[mg/l]

0 ... 20.00
0 ... 90.0

0.01
0.1

Saturation [%]

0 ... 200.0
0 ... 600

0.1
1

D.O. partial pressure
[mbar]

0 ... 200.0
0 ... 1250

0.1
1

T [°C]

0 ... 50.0

0.1