Xylem ProLab 2000 User Manual
Page 64

Operation
ProLab 2000
64
ba75577e05
11/2013
Note
You can terminate prematurely the manual Stability control function
with <MENU/OK> manually at any time. If the manual Stability control
function is terminated prematurely, the current measurement data is
not downloaded to the interface.
Criterion
The meter checks the stability of the measured value on the basis of
the temperature measurement. For identical measurement conditions,
the following applies:
3
With <MENU/OK> activate the Stability control function.
[SC] flashes while the stability control is active.
As soon as a stable measured value is recognized, the current
measurement data is downloaded to the interface. The
measured values of all channels meeting the criterion for
stability control are marked by SC.
4
Using <SC> or <MODE>, release the frozen measured value.
The [SC] status display disappears. The display switches to the
previous indication.
Measured
parameter
Reproducibility
Response time
Temperature
better than 0.02 °C
> 10 seconds