6 ion concentration, 1 general information, Ion concentration 4.6.1 – Xylem ProLab 2000 User Manual

Page 53: General information

Advertising
background image

ProLab 2000

Operation

53

ba75577e05

11/2013

4.6

Ion concentration

4.6.1

General information

Note
Incorrect calibration of ion sensitive electrodes will result in incorrect
measured values. Calibrate regularly before measuring.

CAUTION

When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as incorrect values would result. The
RS232 and USB interfaces are not galvanically isolated.

Temperature

measurement with ISE

measurements

For reproducible measurements of the ion concentration, it is essential
to measure the temperature of the test sample.
You have the following options to measure the temperature:

 Measurement by a temperature sensor.
 Measurement of the temperature by the integrated temperature

sensor of a simultaneously connected D.O. sensor or conductivity
measuring cell in the test sample.

 Manual determination and input of the temperature.

The meter recognizes whether a suitable sensor is connected and
automatically switches on the temperature measurement.

The display of the temperature indicates the active temperature
measuring mode:

* If this is not required you can unplug the second sensor and enter the
temperature manually.

Temperature

sensor

Resolution of the

temp. display

Mode

-

1 °C

Manual

yes

0.1 °C

Automatic with
temperature sensor

yes

0.1 °C,
measured
temperature value
flashes

The temperature value of
the second sensor (Cond
or Oxi) in the same test
sample is taken over for
measurement*