Xylem ProLab 2000 User Manual
Page 77

ProLab 2000
Operation
77
ba75577e05
11/2013
1
Put the D.O. sensor into the air calibration vessel.
2
Connect the D.O. sensor to the meter.
3
In the measured value display, select the D.O. measuring
window with <
▲><▼> and <MODE>.
4
Start the calibration with <CAL>.
The calibration display for the relevant sensor type appears.
5
Put the D.O. sensor into the air calibration vessel.
6
Press <MENU/OK>.
The measurement with stability control to determine the
relative slope starts.
7
Wait for the end of the measurement with Stability control or
accept the calibration value with <MENU/OK>.
The determined relative slope is displayed as a message (
i
).
[i] Relative slope = 1.00
[i] Use air calibration vessel
Continue
LFOx1400-ID <CAL>
[i] Relative slope = 1.00
[i] Temperature =24.8 °C
Terminate stability control
LFOx1400-ID <CAL>