Xylem ProLab 2000 User Manual
Page 70

Operation
ProLab 2000
70
ba75577e05
11/2013
4
Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.
5
Press <MENU/OK>.
The measurement with stability control to determine the cell
constant starts.
6
Wait for the end of the measurement with Stability control or
accept the calibration value with <MENU/OK>.
The determined cell constant is displayed as a message (
i
).
7
You have the following options:
Accept the new calibration values with <MENU/OK>.
Subsequently, the calibration record is displayed and
downloaded to the interface at the same time.
To switch to the measured value display without accepting
the new calibration values, press <MODE> or <ESC>.
[i] Cell constant = 0.475 1/cm
[i] Immerse sensor in solution
Continue
<CAL>
[i] Cell constant = 0.481 1/cm
[i] Temperature =24.8 °C
Terminate stability control
<CAL>
[i] Cell constant = 0.481 1/cm
i Sensor+++
Accept
<CAL>