Xylem ProLab 2000 User Manual

Page 70

Advertising
background image

Operation

ProLab 2000

70

ba75577e05

11/2013

4

Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.

5

Press <MENU/OK>.
The measurement with stability control to determine the cell
constant starts.

6

Wait for the end of the measurement with Stability control or
accept the calibration value with <MENU/OK>.
The determined cell constant is displayed as a message (

i

).

7

You have the following options:
 Accept the new calibration values with <MENU/OK>.

Subsequently, the calibration record is displayed and
downloaded to the interface at the same time.

 To switch to the measured value display without accepting

the new calibration values, press <MODE> or <ESC>.

[i] Cell constant = 0.475 1/cm

[i] Immerse sensor in solution

Continue

<CAL>

[i] Cell constant = 0.481 1/cm

[i] Temperature =24.8 °C

Terminate stability control

<CAL>

[i] Cell constant = 0.481 1/cm
i Sensor+++

Accept

<CAL>