Xylem ProLab 2000 User Manual

Page 69

Advertising
background image

ProLab 2000

Operation

69

ba75577e05

11/2013

Sample record:

Note
If non-ID sensors are used, the sensor type and series number are not
output in the calibration record.

Calibration evaluation

After the calibration, the meter automatically evaluates the current
status of the calibration. The evaluation appears on the display and in
the calibration record.

Determining the cell

constant

Proceed as follows to determine the cell constant:

Printing date

26.04.06 16:13

ProLab 2000
Ser. no.

06249876

USER ID: 1234567890

Calibration Cond
Calibration date 20.04.06 10:14:03
LFOx1400-ID
Ser. no.

A062498765

USER ID: 1234567890

Cal interval

150 d

Cell constant.

0.475 1/cm 25.0 °C

Sensor

+++

Display

Calibration
record

Cell constant [cm

-1

]

+++

within the ranges
0.450 ... 0.500 cm

-1

0.585 ... 0.715 cm

-1

or
0.800 ... 1.200 cm

-1

Error

Error

outside the ranges
0.450 ... 0.500 cm

-1

or
0.800 ... 1.200 cm

-1

Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...

(page 99)

1

Connect a conductivity measuring cell to the meter.

2

In the measured value display, select the conductivity
measuring window with <

▲><▼> and <MODE>.

3

Start the calibration with <CAL>.
The calibration display appears.