Operation

ProLab 2000

76

ba75577e05

11/2013

Sample record:

Note
If non-ID sensors are used, the sensor type and series number are not
output in the calibration record.

Calibration evaluation

After the calibration, the meter evaluates the current status of the
sensor against the relative slope. The evaluation appears on the
display and in the calibration record. The relative slope has no effect on
the measuring accuracy. Low values indicate that the electrolyte will
soon be depleted and the sensor will have to be regenerated.

Calibration in

water vapor

saturated air

(air calibration vessel)

For this calibration procedure, the Comparison meas. setting must be
set to Off in the measuring menu.

Proceed as follows to calibrate the meter:

Note
The sponge in the air calibration vessel must be moist (not wet). Leave
the sensor in the air calibration vessel for a time long enough to adjust.

Printing date 26.04.06 16:13
ProLab 2000
Ser. no.

06249876

USER ID: 1234567890

CALIBRATION Oxi
Calibration date 20.04.06 10:14:03
LFOx1400-ID
Ser. no.

A062498765

USER ID: 1234567890

Cal. interval

14d

Relative slope

0.97

Sensor

+++

Display

Calibration record

Relative slope

+++

S = 0.8 ... 1.25

++

S = 0.7 ... 0.8

+

S = 0.6 ... 0.7

Error

Error

S < 0.6 or S > 1.25

Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

... (page 99)