Xylem ProLab 2000 User Manual
Page 103

ProLab 2000
What to do if...
103
ba75577e05
11/2013
Error message, OFL
Error message, Error
(invalid
calibration)
Error message, LEAK
Measured value flashes
for a long time
(No stable measured
value)
Measured value too low
Measured value too
high
Cause
Remedy
– Measuring range exceeded
– not possible
Cause
Remedy
D.O. sensor:
– Electrolyte solution depleted
– Regenerate probe
– Membrane contaminated
– Clean membrane
– Electrode system poisoned
– Regenerate probe
Cause
Remedy
D.O. sensor:
– Membrane damaged
– Exchange the membrane cap
– Membrane head screwed on
not tight enough
– Screw membrane head tight
Cause
Remedy
D.O. sensor:
– Membrane contaminated
– Clean membrane
Cause
Remedy
D.O. sensor:
– Insufficient flow
– Provide flow to the probe
Cause
Remedy
– High amount of dissolved
substances
– Correct solubility function
using the salinity equivalent
– Air bubbles bump on the
membrane with high velocity
– Avoid direct flow to the
membrane