Technical data

ProLab 2000

108

ba75577e05

11/2013

Stand March 3, 2010

Translation of the legally binding German version

7.2.3

Conductivity

Measuring ranges,

resolution

Variable

Measuring range

Resolution

[µS/cm]

0.000 ... 2.000*
0.00 ... 20.00**
0.0 ... 200.0
0 ... 9999

0.001
0.01
0.1
1

[mS/cm]

0.00 ... 20.00
0.0 ... 200.0
0 ... 9999

0.01
0.1
1

Specific resistance
[kOhm*cm]

0.000 ... 2.000
0.00 ... 20.00
0.0 ... 200.0
0 ... 9999

0.001
0.01
0.1
1

Resistivity
[MOhm*cm]

0.00 ... 20.00
0.0 ... 200.0
0 ... 9999

0.01
0.1
1

SAL

0.0 ... 70.0
according to the IOT
table

0.1

TDS [mg/l]

0 ... 9999
Factor can be set
between 0.40 and
1.00

1

T [°C]

- 5.0 ... + 105.0

0.1

* only possible with cells of the cell constant, 0.010 cm

-1

** only possible with cells of the cell constant, 0.010 cm

-1

or 0.090 ... 0.110 cm

-1

Cell constants

Cell constant C

Values

Can be calibrated in
the ranges

0.450 ... 0.500 cm

-1

0.585 ... 0.715
0.800 ... 1.200 cm

-1

Adjustable

0.010 cm

-1

(fixed)

0.090 ... 0.110 cm

-1

0.250 ... 25.000 cm

-1

Reference temperature

Reference
temperature

Values

Adjustable

20 °C (Tr20)
25 °C (Tr25)