3 temperature compensation, Temperature compensation – Xylem ProLab 2000 User Manual

Page 65

Advertising
background image

ProLab 2000

Operation

65

ba75577e05

11/2013

4.7.3

Temperature compensation

The calculation of the temperature compensation is based on the
preset reference temperature, 20 °C or 25 °C. It appears on the display
as Tr20 or Tr25.

You can select one of the following temperature compensation
methods:

Nonlinear temperature compensation (nLF) according to EN 27

888

Linear temperature compensation (lin) with selectable

coefficients of 0.000 ... 3.000 %/K

 No temperature compensation (off)

Note
The reference temperature and temperature compensation are set in
the conductivity measuring menu (see page 66).

Application tips

Select the following temperature compensations given in the table
according to the respective test sample:

Test sample

Temperature
compensation

Display
indicator

Natural water
(ground water, surface
water, drinking water)

nLF
according to EN 27 888

nLF

Ultrapure water

nLF
according to EN 27 888

nLF

Other aqueous
solutions

lin Set linear
temperature coefficient
0.001 ... 3.000 %/K

lin

Salinity (seawater)

Automatic nLF according to
IOT

Sal, nLF