Step 4 Under the Actions Tab, Select the type of External Trigger

signal suitable for the external instrument. The chassis support
three external trigger-out signal types:

Pulse High

Pulse Low

Pulse Toggle

Note: In each case, the BNC signal transmits an LV TTL level signal

Step 5 Select Trigger Analyzer on the Actions Tab if you intend to

record protocol level traffic simultaneously while using the BNC
External Trigger Out signal

Step 6 Press the Record button to arm the analyzer. Trigger indicator

on the analyzer faceplate will indicate when the trigger condition
has been detected by the analyzer system.

Using the

BNC Trigger

In on a

SASTracer

Step 1 Attach an appropriate BNC Trigger Out connector on a general

purpose test instrument to the BNC Trigger In connector using
the provided BNC cable.

Step 2 In the SASSuite software, under Recording Options, select the

General Tab and click the radio button Event Trigger under
Recording Type.

Step 3 Select the Recording Rules Tab. Remove all triggering events

from the Recording rules (Filtering options can be retained - see
Recording Rules screen image below)