SASSuite User Manual

Version 6.25

LeCroy Corporation

85

When uploading is complete, the index file will be opened. Each "packet" in this file
corresponds to one of the numbered segments. Double clicking on the packet will open
the corresponding segment file.

Multisegment Toolbar

When an index file is opened like the one shown above, the
Multisegment Toolbar will display. This toolbar lets you navigate the
index file.

Spooled Recording

In a Spooled recording, uploading commences when the recording is begun. As traffic is
uploaded from the analyzer to the host PC, the analyzer memory is freed, creating space
for recording additional traffic. Recording can thus continue for long periods of time, and
create file lengths well in excess of the actual physical memory in the analyzer. However,
an issue to be aware of with any spooled recording is that the actual spooling process is
typically slow in comparison to the potential recording speed of the analyzer. So unless
the recorded traffic is intermittent or heavily filtered, it is possible for data to be lost if the
the spooling process does not keep pace with the recording process. This produces
"gaps" in the resulting data, which are identified in the data file.

Open first segment in
multisegment trace.

Open previous
segment in
multisegment trace.

Open index file. This
button becomes active
if a multisegment trace
file is open.

Open next segment in
multisegment trace.

Open last segment in
multisegment trace.