8 error messages (f4-), 8 error messages (f4-) -12 – Yokogawa RAMC User Manual

Page 38

Advertising
background image

<6. ELECTRONIC TRANSMITTER (-E)>

6-12

IM 01R01B02-00E-E 10th edition January 01, 2013 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

6.2.8 Error messages (F4-)

I

f the 8 bars beneath the digits start flashing, an error has occurred in the measuring transmitter/current

output. Since the pointer indication is independent from the electric measuring transmitter, it may show the
correct measuring value even if the transmitter is defective. Function F41 allows checking of assigned error
codes.
Error codes are called onto the display as follows:

Description

Selection

Key

Indication

Change to setting mode

Enter

Display mode
F1-

Setting mode

3 x →

Enter
Enter
Enter

F4-
F41
F41 Enn
F41

Back to display mode

F4-
Display mode

List of error messages :

Code

Meaning

Remedy

01

02
03
04

05
06
07
08

RAM-error
ADC-error
Internal EEPROM faulty
Calibration-EEPROM faulty

Wrong totalizer value in EEPROM
Overflow (flow rate too high)
Internal EEPROM faulty
Float blocking indication realized,
supervision time gone off

Indication unit needs service
Indication unit needs service
Indication unit needs service
If EEPROM is missing insert, -otherwise order new
EEPROM
Reset totalizer
Reduce flow rate
Indicator unit needs service
Deactivate float-blocking-indication or run Autozero
function

In case of error the appropriate remedy has to be taken.