1 process variables, 2 diagnostic- and service-menu, 1 process variables -7 – Yokogawa RAMC User Manual

Page 55: 2 diagnostic- and service-menu -7

Advertising
background image

<7. HART-COMMUNICATION>

7-7

IM 01R01B02-00E-E 10th edition January 01, 2013 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

7.4.1 Process variables

Device setup

Process variables

Flow value [cum/h]

Flow in oriented measurement unit

Flow % rnge [%]

%-value related to 20mA

Flow AO value [mA]

Current output in mA

Totalizer value [kg]

Totalizer

% span value

%-Value related to flow final value

Temperature value [°C]

Temperature in transmitter

7.4.2 Diagnostic- and Service-Menu

Device setup
Diag/Service
Test

device

Device

status

Status

group

1

Error

indication:

RAM Error

OFF/ON Memory error

ADC Error

OFF/ON error A/D converter

ADJ-EE-Error

OFF/ON error adjustment-EEPROM

CAL-EE Error

OFF/ON error calibration-EEPROM

Totalizer false

OFF/ON totalizer value false

Flow Overrun

OFF/ON Flow too high

ADJ Error 2

OFF/ON adjustment-EEPROM fault

Float blocked

OFF/ON Float-Move-Detection recognized, supervision time

passed

Device setup
Diag/Service
Test

device

Device

Status

Status

group

2

Error

indication:

Temp over limit OFF/ON Temperature exceed 70°C

Max flow1 active OFF/ON URV overrun

Max flow2 active OFF/ON URV underrun

FB autoz active OFF/ON Float blocking autozero on

Power fail warn OFF/ON Power off has happened

Oper timer error OFF/ON Operation timer fault

Mancal active

OFF/ON Manual calibration activated

Advertising