<6. ELECTRONIC TRANSMITTER (-E)>

6-15

IM 01R01B02-00E-E 10th edition January 01, 2013 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

For the manual adjustment procedure according to the two cases described the following remarks have to be
taken into account:

After manual adjustment the flow-scale of the indicator is no longer valid.

The display will only indicate in % of the new flow range.

Switching of units is impossible.

The indicator can be always resetted to the original adjustment according to manufacturers

calibration at any time.

The given procedures will only adjust the current output and the display to the new measuring
range for a different fluid and/or new process conditions.

The result of this adjustment is NOT a calibration! If proof of the new adjustment is requested a
real calibration by comparison to a standard has to be made after adjustment!

The following interactions with other functions apply:

Interaction with other functions:

Action

Function

Effect

1. Setting of manual adjustment values

F52 -5...

Manual adjustment table is overwritten

2. Activating of manual adjustment table

F51 -2

- Manual adjustment table active
- %-indication only

- Totalizer does not count
- No other flow rate unit selectable
- Function F64 for option /A2 has no effect,
with manual adjustment

3. Deactivating of manual adjustment

table

F51 -1

- Standard adjustment table active
- F11 is set to flow rate
- Flow unit is the same as before activating
manual adjustment
- Totalizer value is the same as before

activating manual adjustment

Due to the limitations described it is strongly recommended to order a new scale and EEPROM from
manufacturer (this gives a real new flow scale without new calibration) or to order a new calibration by
the manufacturer together with a new scale and EEPROM for the new fluid and/or new process conditions
(this gives new adjustment plus new calibration).