<7. HART-COMMUNICATION>

7-3

IM 01R01B02-00E-E 10th edition January 01, 2013 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

7.3 HART Menu (Rev 01 DD rev 02)

HOT KEY

Write protect

R

Enable wrt 10

min

W

New password

W

Available in generic

mode

R=Read,

W=Write,S=Set

T=Test, p=perform

Device

Process Flow value [cum/h]

R

Setup

Variables Flow % rnge [%]

R

Flow AO value [mA]

R

Totalizer value [kg]

R

% Span value

R

Temperature value

[°C]

R

Flow

Diag/

Test device

Device status

Status group 1

RAM error

R

Flow AO

Service

ADC error

R

ADJ-EE error

R

CAL-EE error

R

Totalizer false

R

Flow LRV

Flow overrun

R

Flow URV

ADJ error 2

R

Float blocked

R

Status group 2

Temp over limit

R

Max flow 1 active

R

Max flow 2 active

R

FB autoz active

R

Power fail warn

R

Oper timer error

R

Mancal active

R

R

Reset all errors

reset

p

Self test

perform

p

Master reset

perform

p

Loop test

4mA/20mA/other/end

T

Calibration

Apply URV/LRV

4 mA

Set as 4mA

value

S

Read new value

S

Leave as found

S

20 mA

Set as 20mA

value

S

Read new value

S

Leave as found

S

Exit

Reset URV/LRV

Reset 4 mA

Perform

p

Reset 20 mA

Perform

p

Exit

D/A trim

4 mA/ 20mA

S

calib table

Std version

R/S

Dist version

R/S

Manual calib

Status mancal

ON / OFF

R

Activate/deactiv

ON / OFF

S

Set mancal points

5%/15%/ ... /95%/105%

S