Yokogawa RAMC User Manual

Page 77

Advertising
background image

<9. TECHNICAL DATA>

9-5

IM 01R01B02-00E-E 10th edition January 01, 2013 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

LIMIT SWITCHES IN STANDARD VERSION

(option /K1 to /K3)

Type :

Inductive proximity switch SC3.5-N0 acc. DIN EN 60947-5-6

Nominal voltage :

8 V DC

Output signal :

≤ 1 mA or ≥ 3 mA

LIMIT SWITCHES IN FAIL SAFE VERSION

(option /K6 to /K10)

Type :

Inductive proximity switch SJ3.5-SN; SJ3.5-S1N

acc. DIN EN 60947-5-6 (NAMUR)

Nominal voltage :

8 V DC

Output signal :

≤ 1 mA or ≥ 3 mA

HYSTERESIS OF LIMIT SWITCHES

Min-contact / Max-contact :

- pointer movement

≈ 0.8 mm

- float movement

≈ 0.8 mm

Minimum distance between 2 contacts :

≈ 2 mm

CABLE GLAND (option /K1 to /K10)

Size :

- M16x1.5 (standard)

- Thread M20x1.5 (option /A13; standard for option /KF1)

- Thread ½” NPT (option /A5)

Cable diameter :

6 – 9 mm

Maximum cross section of core :

Ø 1.5 mm²

POWER SUPPLY FOR LIMIT SWITCHES (Option /W__)

Type :

acc. DIN EN 60947-5-6 (NAMUR)

- KFA5-SR2-Ex*-W (115 V AC), * = 1 or 2

- KFA6-SR2-Ex*-W (230 V AC), * = 1 or 2

- KFD2-SR2-Ex*-W (24 V DC), * = 1 or 2

- KHA6-SH-Ex1 (115/230 V AC), Fail Safe, 1 channel

- KFD2-SH-Ex1 (24 V DC), Fail Safe, 1 channel

Power supply :

- 230 V AC ± 10%, 45-65Hz

- 115 V AC ± 10%, 45-65Hz

- 24 V DC ± 25%

Relay output :

1 or 2 potential-free changeover contact(s)

Switching capacity :

max. 250 V AC, max. 2 A

Note :

If Fail-Safe limit switch option /K6 or /K7 is ordered, for power

supply option /W2E or /W4E must be selected.

If Fail-Safe limit switch option /K8, /K9 or /K10 is ordered, for
power supply option /W2F or /W4F must be selected.

1)

referenced to 20°C ambient temperature

Digital display :

8- digits 7- segment-LC-display character height 6 mm

Process-/ Ambient temperature :

The dependency of the process temperature from the

ambient temperature is shown in fig. 7a to fig. 7c.

The internal temperature of the electronic transmitter

can be indicated on the display or checked via HART

communication.

Measurement of the internal transmitter temperature :

- Range :

-25 °C to +70 °C

- Accuracy : ± 5 °C

Transportation- and Storage condition :

-40 °C to +70 °C

Linearity

1)

:

± 0.2 % f.s.

Hysteresis

1)

:

± 0.1 % f.s.

Repeatability

1)

:

± 0.1 % f.s.

Influence of power supply

1)

:

± 0.1 % f.s.

Temperature coefficient of the output signal

1)

:

± 0.5 % /10 K f.s.

AC-part of output signal

1)

:

± 0.15 % f.s.

Long-time stability

1)

:

± 0.2 % /year

Max. output signal :

21.5 mA

Output signal in case of failure :

≤ 3.6 mA (acc. NE 43)

Response time (99%) :

About 1.5 s (damping 1s)

Electromagnetic compatibility (EMC) :

- Acc. EN 61326-1: 2006, Class A, Table 2 and

EN 61326-2-3 : 2006 :

Criterion A, restriction: HF- immunity between 500 MHz

and 750 MHz : criterion B

RAMC with Profibus PA :

Criterion A: Burst, Surge, HF- Immunity

Criterion B: ESD

In case of single sided grounding of the cable shield it

is possible that for all tests criterion B is reached.

Unit safety acc. DIN EN 61010-1: 2001 :

- Over voltage category : II (acc. VDE 0110 or IEC 664)

- Pollution degree : I

- Safety class :

I (with 115 / 230V AC power supply)

III (with 24V DC power supply and

Fieldbus type)

POWER SUPPLY FOR ELECTRONIC TRANSMITTER

(Option /UT)

Type :

Power supply with galvanically separated input and output

- RN221N-B1, HART- compatible

Supply voltage :

20 ... 250 V DC / AC 50/60 Hz

Maximum load :

700 Ω

Output signal :

4 - 20 mA

CABLE GLAND (for transmitter –E, -H, -J and –G) :

Size :

- M16x1.5 (standard)

- Thread M20x1.5 (option /A13; standard for option /KF1)

- Thread ½” NPT (option /A5)

Cable diameter :

6 – 9 mm

Maximum cross section of core :

Ø 1.5 mm²