Yokogawa RAMC User Manual

Page 40

Advertising
background image

<6. ELECTRONIC TRANSMITTER (-E)>

6-14

IM 01R01B02-00E-E 10th edition January 01, 2013 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

Input of manual adjustment table (F52)
The manual adjustment table is input as follows:

Description

Selection

Key

Indication

Change to setting mode

Enter

Display mode
F1-

Setting mode

4 x →

Enter

F5-
F51
F52

Selection

5%- point
15%- point
25%- point
35%- point
45%- point
55%- point
65%- point
75%- point
85%- point
95%- point
105%- point

Enter
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Enter

F52
F52 -5
F52 -15
F52 -25
F52 -35
F52 -45
F52 -55
F52 -65
F52 -75
F52 -85
F52 -95
F52 -105

Back to display mode

F5-
Display mode

2 Manual “wet” adjustment by comparison to a reference master meter with the real process fluid at

operating conditions:

This adjustment is useful under the following conditions:

The original manufacturer’s calibration is not available or needs to be renewed.

Or

The user is not able to recalculate the new mm to flow rate table.

And

The user has the possibility to compare the meter indication with a master meter with the

process fluid at process conditions.

In these cases the following steps have to be performed:

2.a ) Place the RAMC in line with the master meter in an installation allowing controlled flow with the

process fluid at process conditions in a flow range from 5% to 105% of the expected flow range.

2.b ) Go to menu function F51 and press ENTER to switch to manual adjustment mode. (Switching back

to the original adjustment is possible by pressing ENTER again).

2.c ) Go to menu function F52 in order to start the manual adjustment.
2.d ) Set the flow to 5% of the new flow rate indicated by the master.
2.e ) Press ENTER to adjust the first 5% point.
2.f ) Repeat steps 2.d) to 2.e) for the 15%; 25%; 35%; 45%; 55%; 65%; 75%; 85%; 95% and 105%

points. (Note: The whole loop from 5% to 105% has to be adjusted in the requested order without

interruption. It is not possible to skip or stop and restart the adjustment.)

2.g ) The adjustment has to be finished and stored by pressing “ ↑ “.

After storage the new adjustment is permanently available and can be switched “on” or “off” by function
F51.