Warning, Tag plate for intrinsically safe type, 9-32 nakacho, musashino-shi, tokyo japan – Yokogawa Wireless Temperature Transmitter YTA510 User Manual

Page 15

Advertising
background image

<2. Notes on Handling>

2-7

IM 01C50E01-01EN

Temperature

Transmitter

Battery

pack

1
2
3
4
5

1
2
3
4
5

Temperature

Transmitter

Battery

pack

Temperature Sensor (*1)

Amplifier housing code 8 and 9

Hazardous area

Antenna

(*1)

Arrester

(*1, *2)

Temperature Sensor (*1)

(One or Two sensors)

*1: These apparatus are simple apparatus.

*2: Arrester may not be connected.

Antenna Connector

F0206.ai

[Control Drawing]

Amplifier housing code 7

Hazardous area

Note 4. Battery Pack

• Use only YOKOGAWA battery pack

F9915MA or F9915NS.

Note 5. Special conditions for Safe Use

• In case of the aluminum enclosure, if the

Transmitter is mounted in an area where the

use of category 1 G apparatus is required,

it must be installed such, that, even in the

event rare incidents, ingition source due to

impact and friction sparks are excluded.

• For application in an area where the use

of category 1 G apparatus is required,

electrostatic charge on the non-metallic

parts of the Temerature Transmitter shall be

avoided

WARNING

Potential electrostatic charging hazard - secure

distance of 100mm from antenna.

(2) Operation

WARNING

Take care not to generate mechanical sparking

when access to the instrument and peripheral

devices in a hazardous location.

(3) Maintenance and repair

WARNING

The instrument modification or parts replacement

by other than an authorized Representative of

Yokogawa Electric Corporation is prohibited and

will void the certification.

(4) Name Plate

• Name Plate

: Refer to USER'S MANUAL.

Made in Japan

TOKYO 180-8750 JAPAN

MODEL

SUFFIX

SUPPLY

OUTPUT

MWP

mA DC

V DC

STYLE

CAL

RNG

NO.

F0207.ai

TEMPERATURE

TRANSMITTER

• Tag plate for intrinsically safe type

F0208.ai

POTENTIAL ELECTROSTATIC CHARGING HAZARD - SECURE DISTANCE

OF 100MM FROM ANTENNA.

USE ONLY BATTERY PACK YOKOGAWA F9915MA OR F9915NS.

POTENTIAL ELECTROSTATIC CHARGING HAZARD - SEE USER'S MANUAL.

WARNING

No. KEMA 10ATEX0163 X

Ex ia IIC T4 Ga

ENCLOSURE: IP66/IP67

Tamb.: -50 TO 70°C

Sensor input circuit (Terminal 1 to 5)

Uo=6.6V, Io=66mA, Po=109mW, Co=22μF, Lo=8.1mH

*3

MODEL: Specified model code.

SUFFIX: Specified suffix code.

STYLE: Style code.

SUPPLY: Supply voltage.

OUTPUT: Output signal.

CAL RNG: Calibration range.

NO.: Serial number and year of production

*1

.

TOKYO 180-8750 JAPAN: The manufacturer

name and the address

*2

.

*1:

The first digit in the final three numbers of the serial

number appearing after “NO.” on the nameplate indicates

the year of production. The following is an example of a

serial number for a product that was produced in 2010:

91K819857 032

The year 2010

*2:

“180-8750” is a zip code which represents the following

address.

2-9-32 Nakacho, Musashino-shi, Tokyo Japan

*3:

The identification number of Notified Body.