1 replacing the integral indicator, 2 replacing the rf assembly, Replacing the integral indicator -3 – Yokogawa Wireless Temperature Transmitter YTA510 User Manual

Page 54: Replacing the rf assembly -3

Advertising
background image

<8. Maintenance>

8-3

IM 01C50E01-01EN

RF assembly

CPU assembly

Amplifier Cover

Mounting screw

Integral

Indicator

Boss

Stud

F0802.ai

Press forward

Output terminal

cable

Figure 8.2

Removing and Attaching integral

Indicator and CPU Assembly

8.4.2 Replacing the RF Assembly

This subsection describes how to replace the RF

assembly (see Figure 8.2).
■ Removing the RF assembly
1) Remove the cover.

2) Remove the Integral indicator (refer to

subsection 8.4.1).

3) Remove the two stud bolts by using a socket

driver (width across flats: 5.5 mm)

4) Disconnect the RF assembly from the CPU

assembly. When doing this, carefully pull the RF

assembly straight forward so as not to damage

the connector pins between it and the CPU

assembly.

5) Disconnect the antenna cable that connects the

CPU assembly and the antenna.

NOTE

Be careful not to apply excessive force to the

RF assembly and the connector of the antenna

cable when removing it.

■ Mounting the RF assembly
1) Connect the antenna cable between the RF

assembly and the antenna.

2) Align both the RF assembly and CPU assembly

connectors and engage them.

3) Tighten the two stud bolts.

4) Mount the Integral indicator (refer to subsection

8.4.1).

5) Replace the cover.

Table 8.2

Tools for Disassembly and Reassembly

Tool

Quantity

Remarks

Phillips

screwdriver

1

JIS B4633, No. 2

Slotted

screwdriver

1

Allen wrenches

3

JIS B4648

One each, nominal 3,

4 and 2.5 mm Allen

wrenches

Wrench

1

Width across flats, 17 mm

Torque wrench

1

Adjustable

wrench

1

Socket wrench

1

Width across flats, 16 mm

Socket driver

1

Width across flats, 5.5 mm

Tweezers

1

8.4.1 Replacing the Integral Indicator

This subsection describes the procedure for

replacing an integral indicator. (See figure 8.2)

■ Removing the Integral Indicator
1) Remove the cover.

2) While supporting the integral indicator with one

hand, loosen its two mounting screws.

3) Dismount the LCD board assembly from the RF

assembly.

When doing this, carefully pull the LCD board

assembly straight forward so as not to damage

the connector pins between it and the RF

assembly.

■ Attaching the Integral Indicator
1) Align both the LCD board assembly and RF

assembly connectors and engage them.

2) Insert and tighten the two mounting screws.

3) Replace the cover.