<10. General Specifications>

10-1

IM 01C50E01-01EN

10. General Specifications

10.1 General Specification

Communication Protocol

ISA100.11a protocol

Data Rate

250 kbps

Frequency

2400 - 2483.5 MHz license free ISM band

Radio Security

AES 128 bit codified

RF Transmitter Power

Max. 11.6 dBm (fixed)

Antenna

+2 dBi Omni directional monopole type

Measurement Range

See Table 10.1

Accuracy

See Table 10.1

Cold Junction Compensation Accuracy

± 0.5°C

Ambient Temperature Effect

See Table 10.2

Battery Pack

Refer to GS 01C50E01-01EN

Update Time

1 to 3600 s selectable*

* Minimum update time is 2 s at dual input sensor use.

Input

Single input: Amplifier housing code 7

Dual input: Amplifier housing code 8 and 9

Input type is selectable: Thermocouples, 2-, 3-,

and 4-wire RTDs, ohms and DC milivolts. See

Table 10.1.

4-wire is available for Sensor 1 input. See Table

10.1.

Input Signal Source Resistance ( for T/C, mV )

1 kΩ or lower

Input Lead Wire Resistance ( for RTD, Ohm )

10 Ω per wire or lower

Power Supply Specifications

Battery:

Use the dedicated battery pack.

Rated voltage: 7.2 V

Rated capacity: 19 Ah

Ambient Temperature Limits

- 40 to 85°C ( - 40 to 185°F )

- 30 to 80°C ( - 22 to 176°F ) LCD visible range

Ambient Humidity Limits

0 to 100% RH

Zero-gain Adjustment

Set the amount of zero-gain point adjustment

Integral Indicator (LCD display)

5-digit numerical display, unit display and bar

graph. The indicator is configurable to display

the following variables periodically.

°C, K, °F, °R, mV and Ω, 0 to 100 % bar graph,

and alternate display of Sensor 1 and Sensor 2

process value

Sensor Burnout

Select either HIGH or LOW as the

configuration.

Self Diagnostics

Amplifier failure, sensor failure, configuration

error, battery alarm, wireless communication

alarm and over-range error for process

variables.

EMC Conformity Standards

EN61326-1 Class A, Table 2 (For use in

industrial locations), EN61326-2-3

Advertising