<7. Setting Parameters>

7-15

IM 01C50E01-01EN

- Setting to enable or disable the hardware write

protection switch.

- Setting to enable or disable changing the

setting to the Diagnostic Switch and Diagnostic

Configuration parameters.

2) Procedure to call up the protection setting

display parameter (Hardware Write Protect)

UAPMO block : Hardware Write Protect
The Hardware Write Protect parameter enables

the switch status of hardware write protection to

be displayed.

For the relationship between hardware write

protection and software write protection, refer to

section 9. "Parameter Summary".

7.3.13 Input Calibration

The input calibration function is used to adjust

to the transmitters internal interpretation of the

input signal. Perform the calibration procedure for

Sensor1 and Sensor2 separately.

Two points, namely, the lower adjustment point

called Zero point and the upper adjustment point

called Gain, are adjusted, respectively. There are

two adjustment methods. One method uses a

reference signal generator (e.g. a voltage generator

or variable resistor) and the other uses the signal

from a temperature sensor immersed in the solution

with a known temperature.

Input adjustment is performed by inputting (writing)

the signals (voltage, and resistance values)

currently being input as the parameters. This

instrument automatically corrects the error between

the written data and the internally calculated value.

Follow the procedure below for Sensor1 (AI1) for

example.
1) Setting the Input Adjustment Value

a) Setting the AI1 block write mode
• AI1 block : Mode : Target
Set the Target parameter to O/S.
b) Setting the update period
• CO block: Data publication period
Set the data publication period parameter

to 1 second. The computation period in the

transmitter will also be synchronized to the

update period and set to 1 second.

c) Setting the sensor type
• AI1 block: Lin Type
Set the sensor type used in the Lin Type

parameter.

d) Setting the cold conjunction compensation
• AI1 block: Selection of Reference Junction
Set the Selection of Reference Junction

parameter to No Reference (0) in order

to remove the effect of the cold junction

compensation when using the thermocouple for

the sensor type.

e) Perform Zero Point Adjustment using a

reference input generator. The input value

is adjusted near the zero point within the

measurement range used.

f) Set the Zero Point Adjustment value when the

input is stabilized.

• AI1 block: Cal Point Lo

Set the Cal Point Lo parameter to the voltage

value (unit: mV) when the temperature sensor

is the thermocouple, and set the resistance

value (unit: Ω) when the temperature sensor is

the resistance temperature sensor.

g) Perform gain-point adjustment using a

reference input generator. The input value

is adjusted near the gain point within the

measurement range used.

h) Set the gain-point adjustment value when the

input is stabilized.

• AI1 block: Cal Point Hi
Set the Cal Point Hi parameter to the voltage

value (unit: mV) when the temperature sensor

is the thermocouple, and set the resistance

value (unit: Ω) when the temperature sensor is

the resistance temperature sensor.

i) Setting the AI1 block write mode
• AI1 block : Mode : Target
Set the Target parameter to Auto and ensure

that the setting error of the adjustment value

has not occurred.