3-2. sensor cable connection using terminal box, 4. ph201g*b dedicated distributor, 4-1. communication setup – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 102: 3-2. sensor cable connection using terminal box -5, 4. ph201g*b dedicated distributor -5, 4-1. communication setup -5

Advertising
background image

IM 12B07D02-01E

Appendix 11-5

When the option of the screw terminal "/TB" is specified, the terminal block layout is as follows.

+

G

11

14

13 NC

HIGH IMP 16

NC

12

17 LOW IMP

NC

15

Note: Nothing is connected to NC

F3.8E.eps

When a pH sensor is used, connect both terminals LOW and IMP on the upper row with a shorting

bar and nothing to both terminals HIGH and IMP on the lower row.

When an ORP sensor is used, connect both terminals LOW and IMP on the upper row with a shorting

bar as well both terminals HIGH and IMP on the lower row with another one.

11-3-2. Sensor cable connection using terminal box

If you can't install a sensor close to the PH202 transmitter, you may need to connect to an

intermediate terminal box (WTB10-PH1, -PH2) and use an extension cable. You should order these

from Yokogawa, as these Yokogawa parts are specially designed so that system specifications will be

met. The total cable length (for 5 m electrode cable plus 15 m extension cable) may be up to 20 m.

Refer to Figure 11-1 for the wiring.

11-4. PH201G*B Dedicated Distributor

The PH202 supports proprietary bidirectional serial communications with the PH201G (Style B) for

remote maintenance, washing and remote diagnostic purposes. This allows PH202 status (under

maintenance/abnormal) to be monitored remotely, and commands to be sent to an instrument for

wash. Prepare for the instrument separately. Service codes 60 relates to communication settings.

11-4-1. Communication setup

Code 60 *COMM. The settings should be adjusted to suit the communicating device connected to the

output. The communication can be set to HART or to PH201G (Style B) distributor.

When used with our PH201G (Style B) you can enable or disable contact outputs,

namely, Fail contact and Hold contact. The PH201G (Style B) can output Hold

contact and Fail contact signals. You can set Service Code 53 to “0” for “soft fail”

to

disable.

Fail contact output. When you set Service Code 53 to “1” for “hard fail”, set Service

Code 60 to “2.0” to enable Fail contact output of PH201G (Style B), or set Service

Code 60 to “0.1” to disable Fail contact output of PH201G (Style B).

*ADDR. Select address 00 for point communication with 4-20 mA transmission. Addresses

01 to 15 are used in multi-drop configuration (fixed 4 mA output).

Code Display Function

Function detail

X

Y

Z Ddefault values

Communication
60

*COMM Communication

Set communication Off

0

1.0(

*

) Off

Set communication On

1

write

enable

0

write

protect

1

Set communication PH201G*B

2

Without half time check by setting 2.0

0

With half time check by setting 2.1

1

*ADDR. Network address

Set address 00 to 15

00

T5.3.6.eps

(*) Valid for instrument using software version 3.1 or later.
For earlier versions the default is 0.1.