7-2 maintenance

7-3. Calibration procedures are described in step-by-step detail in chapter 6. However, follow these

guidelines.

1.

Before starting a calibration, make sure the electrode system is properly cleaned so that electrodes

are fully functional. They must then be rinsed with clean water to avoid contamination of the

calibration solution.

2.

Always use fresh buffer solutions to avoid the possibility of introducing errors from contaminated or

aged solutions. Buffers supplied as liquids have a limited shelf life, especially alkaline buffers which

absorb CO

2

from the air.

3.

Yokogawa strongly recommends NIST (primary) buffer standards in order to ensure the best

accuracy and best buffer capacity is available. Commercially adjusted buffers (e.g. 7.00, 9.00

or 10.00 pH) are a compromise as a standard, and are often supplied without the temperature

dependency curve. Their stability will be much worse than for NIST solutions.

NOTE

:

NIST (formerly NBS) buffers are available as consumable items from any Yokogawa sales office

under the following part numbers:

6C232 4.01 pH at 25°C}

6C237 6.87 pH at 25°C} A box contains 5 packets of powder. Each makes a 200 ml solution.

6C236 9.18 pH at 25°C}