3-8 Installation and wiring

Figure 3-9. Terminal identification labels

3-6. Sensor wiring

Refer to figure 3-10, which includes drawings that outline sensor wiring.

For sensor wiring of PH8EFP, PH8ERP, PH8EHP, OR8EFG, OR8ERG, HA405, HA406, HA485,

DPA405, DPA406, DPA485, HF405, DPAS405, DPAS485 and FU20, see APPENDIX 2 (Chapter 11).

The EXA analyzers can be used with a wide range of commercially available sensor types, both from

Yokogawa and other manufacturers. The sensor systems from Yokogawa fall into two categories; the

ones that use a fixed cable and the ones with separate cables.

To connect sensors with fixed cables, simply match the terminal numbers in the instrument with the iden-

tification numbers in the instrument on the cable ends.

The separate sensors and cables are not numbered, but instead use a color-coding system. The elec-

trodes have a colored band incorporated in the label on the connection cap:

• Red

for measuring electrodes (both pH and ORP)

• Yellow

for reference electrodes

• Blue

for combined sensors with both measuring and reference elements in the same body

• Green

for temperature sensors

The recommended procedure is to color-code each end of the cables to match the sensors with the color

strips provided with each cable. This provides a quick way to identify the ends of the cables belonging to

a particular sensor when they are installed. (The procedure for fixing the identification labels is described

in detail in the instruction sheet provided with the cable.)