Ma ma ma – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 85

Advertising
background image

IM 12B07D02-01E

Troubleshooting 8-3

Table 8-1. Error Codes

Code Error description

Possible cause

Suggested remedy

E0

Buffer solution temperature

Buffer solution too hot or too cold

Adjust buffer temperature

outside the programmed range

Check cabling

E1

Measurement failed to stabilize.

Sensors fouled

Clean sensors

during the calibration

Sensors too slow (aged sensor)

Replace sensors

E2

Asymmetry potential too high.

Sensors are aged or polluted

Check buffer solution

(Limits set in service code 21.)

Mistake in calibration

Recalibrate at pH7

Replace

sensor

E3

Slope (sensitivity) is outside limits.

Measuring sensor aged

Replace measuring sensor

(Limits set in service code 22.)

Poor insulation at the connector

Replace or dry cables

E4.1 Impedance of input 1 too low.

Measuring sensor broken

Replace measuring sensor

(Limits set in service code 03.)

Damaged or damp connections

Replace or dry cable

E4.2 Impedance of input 2 too low.

Reference sensor broken

Replace reference sensor

(Limits set in service code 04.)

Damaged connections

Replace cables

E5.1 Impedance of input 1 too high.

Measuring sensor disconnected

Check connections

(Limits set in service code 03.)

Sensors not immersed in process

Check process

Liquid earth disconnected

Check connections

E5.2 Impedance of input 2 too high.

Reference sensor fouled

Clean or replace sensor

(Limits set in service code 04.)

Liquid earth disconnected

Check sensor immersion

Insufficient electrolyte

Check electrolyte reservoir

E7

Temperature sensor open

Process too hot or too cold

Check process

> 140°C (or <-10°C for 8k55)

Wrong temperature sensor setting

Check sensor & setting

Temperature sensor damaged

Check connections

E8

Temperature sensor shortened

Process too cold or too hot

Check process

< -30 °C (or > 120 °C for 8k55)

Wrong temperature sensor used

Check sensor & setting

Temperature sensor damaged

Check connections

E9

Measurement out of range (-2 to 16 pH)

Sensors disconnected

Check cabling

Sensor wrongly connected

Check cabling

Sensor(s) defective

Replace sensor(s)

E10 EEPROM write failure

Fault in electronics

Try again, if unsuccessful

contact

Yokogawa

E11 Wash recovery check error

Measuring sensor aged

Replace measuring sensor

(if communication is set to pH201*B in code

Sensor still coated after washing

Check cleaning system

60)

Defective wash system

If needed adjust timings

E12 ORP / rH outside of preset limits

Sensors disconnected or wrongly

Check cabling

connected

E14 No valid calibration data.

Data lost after switching from

Recalibrate

pH to ORP

E15 Cable resistance to temperature

Cable resistance too high

Use Pt1000Ω

sensor exceeds limit value.

Corroded contacts

Clean and reterminate

Wrong sensor programmed

Reprogram

E16 Call for maintenance interval time exceeded. System not maintained in

Perform maintenance

preset time period

Reset interval

E17 Output span too small < 1pH

Incorrect configuration by user

Reprogram

E18 Table values make no sense
E19 Programmed values outside acceptable limits Incorrect configuration by user

Reprogram

E20 All programmed data lost

Fault in electronics

Contact Yokogawa

Very severe interference

E21 Checksum error

Software problem

Contact Yokogawa

E23 Zero point outside limits

Sensors are aged or polluted

Check buffer solution

Mistake in calibration

Recalibrate at pH7

Replace

sensor

mA

mA
mA

Advertising