IM 12B07D02-01E

Installation and wiring 3-7

3-5. Wiring the sensor system

3-5-1. Impedance measurement jumper settings

NOTE:

It is important to decide first which application and which settings are appropriate for the

installation. This decision is best made before the jumpers are installed, because the cables will

rest beside the jumpers in their installed positions.

Table 3-1. Impedance measuring jumpers

Figure no.

Jumper Settings

Jumper Settings

Application & Sensor Connections

Input #1

Input #2

1

High Impedance

Low Impedance

Normal pH sensors

Glass

sensor

on

Input

#1

Reference

sensor

on

Input

#2

2

High Impedance

High Impedance

Special electrodes using

2

glass

sensors

(e.g.

Pfaudler

18)

3

Low Impedance

High Impedance

ORP (pH compensated) and/or rH

metal

sensor

on

Input

#1

pH

glass

(as

reference)

on

Input

#2

4

Low Impedance

Low Impedance

ORP (Redox measurement)

metal

sensor

on

Input

#1

Normal

reference

on

Input

#2

For convenience insulated jumper links are provided. Ordinary wire links can also be used, and are just

as effective.

The following four jumper figure illustrations (figure 3-8) show the jumper positions related to the figure

numbers in the above table.

11 12 14 17 13

15 16

HIGH IMP

LOW IMP

SUPPLY

TEMP LE

INPUT

2

INPUT

1

11 12 14 17 13

15 16

HIGH IMP

LOW IMP

SUPPLY

TEMP LE

INPUT

2

INPUT

1

11 12 14 17 13

15 16

HIGH IMP

LOW IMP

SUPPLY

TEMP LE

INPUT

2

INPUT

1

11 12 14 17 13

15 16

HIGH IMP

LOW IMP

SUPPLY

TEMP LE

INPUT

2

INPUT

1

1

2

3

4

Setting on shipment