Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 25

Advertising
background image

IM 12B07D02-01E

2-12 Specification

2-9. Control Drawing of PH202S FF/PB Specification (IECEx)

Safe area

Hazardous area

Zone 0 or 1

Safe area

Apparatus

I.S.

interface

I.S.

certified

Terminator

Ui = 24 V

or Ui = 17,5 V

Ii = 250 mA Ii = 380 mA

Pi = 1,2 W Pi = 5,32 W

Ex ia IIC

T4 for ambient temp. d 55 qC

PH202S-F

or PH202S-P

I.S.

certified

Terminator

-

+

-

+

Transmitter

Transmitter

Sensor

Connections

-

+

-

+

x

Sensor(s) are of a passive type to be regarded as 'simple apparatus'.

x

Electrical data of the PH202S-F & PH202S-P:

-

Supply and output circuit:

Maximum input voltage Ui=24 V

Maximum input current Ii=250 mA

Maximum input power Pi=1.2 W

Effective internal capacitance Ci= 220 pF;

Effective internal inductance Li= 0

μH.

or

FISCO field device

Maximum input voltage Ui=17.5 V

Maximum input current Ii=380 mA

Maximum

input

power

Pi=5.32

W

Effective internal capacitance Ci= 220 pF;

Effective internal inductance Li= 0

μH.

-

Sensor input circuit:

Maximum output voltage Uo=14.4 V; Maximum output current Io=32.3 mA

Maximum allowed external capacitance Co=600 nF

Maximum allowed external inductance Lo=34 mH

x

Any I.S. interface may be used that meets the following requirements:

FISCO power supply

Uo d 24 V or Uo d 17.5 V

Io d 250 mA Io d 380 mA

Po d 1.2 W Po d 5.32 W

Co t 220 pF + Ccable; Lo t 0

μH + Lcable

x

Electrical data of the PH202S-B & PH202S-D (Type of protection “n”)

-

Supply and output circuit:

Maximum input voltage Ui = 32 V

Effective internal capacitance Ci = 220 pF; Effective internal inductance Li = 0

μH.

-

Sensor input circuit:

Maximum output voltage Uo=14.4 V; Maximum output current Io=32.3 mA

Maximum allowed external capacitance Co = 3.5

μF

Maximum allowed external inductance Lo = 76 mH