2. preparation, 2-1. cables, terminals and glands – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 36

Advertising
background image

IM 12B07D02-01E

Installation and wiring 3-3

3-2. Preparation

Refer to figure 3-4. The power/output connections and the sensor connections should be made in

accordance with the diagram on page 3-6. The terminals are of a plug in style for ease of mounting.

To open the EXA 202 for wiring:

1. Loosen the four frontplate screws and remove the cover.

2. The terminal strip is now visible.

3.

Connect the power supply. Use the gland on the left for this cable.

4.

Connect the sensor input, using the gland on the right (see fig. 3-5). Switch on the power.

Commission the instrument as required or use the default settings.

5. Replace the cover and secure frontplate with the four screws.

6. Connect the grounding terminals tp protective earth.

7.

The optional hose connection is used to guide the cables coming from an immersion fitting through a

protective plastic tubing to the transmitter.

3-2-1. Cables, terminals and glands

The PH202 is equipped with terminals suitable for the connection of finished cables in the size range:

0.13 to 2.5 mm (26 to 14 AWG). The glands will form a tight seal on cables with an outside diameter in

the range of 6 to 12 mm (0.24 to 0.47 inches).

Figure 3-5. Glands to be used for cabling

Grounding

terminal

Power/Output

cable gland

Sensor

cable gland