IM12B07D02-01E

5-12 Parameter setting

Code Display

Function

Function

detail

X

Y

Z

Default

values

Parameter specific functions
01

*PH.ORP

Select main parameter pH

0

0

pH

ORP

1

02

*PRM.2

Enable 2nd parameter Off

0

0

Off

ORP

1

rH

2

03

*Z1.CHK

Impedance check 1

Low

0

1.1.1

High

1 High

Temp comp off

0

Temp comp on

1

On

Imp check off

0

Imp check on

1

On

*Z.L.xΩ

Low impedance limit Press NO to step through choice of units,

1

x = None, K, M or G press YES to select units, then use the

>, ^ ENT keys to set the value

*Z.H.xΩ

High impedance limit Press NO to step through choice of units,

1

x=None, K, M or G

press YES to select units, then use the

>, ^ ENT keys to set the value

04

*Z2.CHK

Impedance check 2

Low

0

0.0.1 Low

High

1

Temp comp off

0

Off

Temp comp on

1

Imp check off

0

Imp check on

1

On

*Z.L.xΩ

Low impedance limit Press NO to step through choice of units,

100

Ω

x = None, K, M or G press YES to select units, then use the

>, ^ ENT keys to set the value

*Z.H.xΩ

High impedance limit Press NO to step through choice of units,

200

x = None, K, M or G press YES to select units, then use the

>, ^ ENT keys to set the value

05

*CAL.CK

Calibration check

Asymmetry check off

0

1.1

Asymmetry check on

1

On

Slope check off

0

Slope check on

1

On

06-09

Not

used