6-3. supplement of temperature sensor, 6-4. auto-return, 6-5. calend – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 112: 6-6. how to cancel sample calibration, 6-3. supplement of temperature sensor -15, 6-4. auto-return -15, 6-5. calend -15, 6-6. how to cancel sample calibration -15, Caution

Advertising
background image

IM 12B07D02-01E

Appendix 11-15

When the OR8EFG or OR8ERG sensor is used, set both INPUT1 and INPUT2 to the low impedance

measurement. Confirm the position of the jumper cable of the wiring diagram.

11 12 14 17 13

15 16

HIGH IMP

LOW IMP

SUPPLY

TEMP LE

INPUT 2

INPUT 1

Setting for ORP

The temperature compensation of the impedance measurement is for conventional pH glass sensors.

When other sensors are used, switch this feature off.

11-6-3. Supplement of temperature sensor

When the temperature sensor is not connected with the transmitter (when no connections are made

to terminals 11 and 12), set the manual temperature compensation (refer to Section 5-1-1).

Yokogawa’s pH sensors, Models HA405, DPA405, DPAS405 and HF405, do not have a built-in

temperature sensor.

11-6-4. Auto-Return

Manual HOLD is automatically turned OFF.

Calibration mode or Commissioning mode is changed to Measurement mode automatically.

If no key is operated for 10 minutes in any mode other than measurement mode or after 10 minutes

in Hold status, Auto-Return (factory setting: On (1) in service code 50) will be activated to return the

transmitter to measurement mode. To disable Auto-Return, set the service code 50 to Off (0) .

If you disable Auto-Return function then the transmitter does not automatically return to Measurement

Mode. You need to press [MODE] key to return to Measurement mode.

11-6-5. CALEND

After CAL.END appears once during calibration, it blinks and calibration starts again.

For the PH202, even after CAL.END is displayed, a measurement stability check continues for the

time interval set by service code 20. If the reading changes significantly* during this interval*, then

recalibration will start automatically. In auto-calibration mode (AUT.CAL) immediately after CAL.END

is displayed, you can press YES or NO within time interval set by service code 20 to confirm or reject

(repeat) the reading.

This applies to both calibration points.
1. If the reading does not stabilize (and CAL.END appear) within 10 minutes of start of calibration,

then error E1 (measurement unstable during calibration) is generated.

2. If you press YES or NO then you can continue immediately to the next step.
3. If you don't press YES or NO and the pH is stable (within (pH) for time interval (t)), then calibration

continues to the next step.

11-6-6. How to cancel sample calibration

T03.ai

In case of stopping calibration at data aquisition situation

(waiting for measured value input with blinking "SAMPLE")

after sample calibration start, cancel the calibration in accordance

with the right table.

Indication on LCD

"SAMPLE" blink

"LAB.CAL" light

"SAMPLE" light

"S. ON" light

"S.OFF" light

Key Operation

Push "MODE" key

Push "NO" key

Push "YES" key

Push "NO" key

Push "YES" key