5. calibration procedures, 5-1. automatic calibration, 5. calibration procedures -2 – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 75: 5-1. automatic calibration -2

Advertising
background image

IM 12B07D02-01E

6-2 Calibration

pH

pH

YES

NO

YES

NO

YES

NO

pH

YES

NO

pH

NO

pH

YES

Press YES for single point

(As Pot) adjustment.

Press NO to proceed to calibrate Slope.

The instruments waits for

the reading to stabilize.

(The pH display flashes)

When reading Is stable,

the CAL END message

appears.

Transfer to second buffer (pH4) and press YES.

After briefly displaying WAIT, the display

returns to the normal readout.

The display now flashes the

pH value. The instrument

waits for the reading to stabilize.

NO

YES

pH

NO

YES

pH

YES

pH

ENT

MAN.CAL

DISPLAY

HOLD

NO

MODE

YES

YOKOGAWA

MODE

TEMP

AUT.CAL

MEASURE

NO

YES

pH

Automatic Calibration

Press the MODE key. AUT.CAL appears in the display, and the

YES/NO

key prompt flags flash. Press YES.

NEW.SNS: Display flashes YES/NO

answer YES if new sensor fitted or NO if not.

Care! YES resets logbook calibration data.

Insert the sensors in pH 7 buffer solution.

Press YES to start calibration.

NO

Note: To start calibration with another solution, press NO to cycle through the other options.

"CAL 7" NO "CAL 4" NO "CAL 9" NO "CAL 7"

YES

When the reading is stable, the CAL END

message appears. Press YES.

After briefly displaying WAIT, the display returns to

the normal readout.

6-5. Calibration procedures

6-5-1. Automatic calibration