Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 73

Advertising
background image

IM12B07D02-01E

5-22 Parameter setting

Code Display

Function

Function

detail

X

Y

Z

Default

values

Communication
60

*COMM.

Communication

Set HART® communication

Off

0

1.0

On

Set HART® communication

On

1

write enable

0

write

write

protect

1

enable

Set communication PH201

*

B

2

Without half time check

0

With half time check

1

*ADDR.

Network address

Set address 00 to 15

00

61

*HOUR

Clock setup

Adjust to current date and time using

*MINUT

>, ^ and ENT keys

*SECND
*YEAR
*MONTH
*DAY
62

*ERASE

Erase logbook

Press YES to clear logbook data

63-69

Not

used

mA

mA

Code Display

Function

Function

detail

X

Y

Z

Default

values

General
70

*LOAD

Load defaults

Reset configuration to default values

71-78

Not

used

79

*CUST.D

Load Customer

Reset configuration to default values

Defaults

except buffer tables

Code Display

Function

Function

detail

X

Y

Z

Default

values

Test and setup mode
80

*TEST

Test and setup

Not used