3-6. communication setup, 3-7. general, 3-8. test and setup mode – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 72: 3-6. communication setup -21, 3-7. general -21, 3-8. test and setup mode -21

Advertising
background image

IM12B07D02-01E

Parameter setting 5-21

5-3-6. Communication setup

Code 60

*COMM.

The settings should be adjusted to suit the communicating device connected to

the output. The communication can be set to HART® or to PH201*B distributor

(for Japanese market only)

*ADDR.

Select adress 00 for point to point communication with 4-20 mA transmission.

Address 01 to 15 are used in multi-drop configuration (fixed 4 mA output).

For the Yokogawa PC202 software package, the default settings match the

software as shipped.

Code 61

*HOUR

The clock/calendar for the logbook is set for current date and time as

reference.

*MINUT

*SECND

*YEAR

*MONTH

*DAY

Code 62

*ERASE

Erase logbook function to clear the recorded data for a fresh start. This may

be desirable when re-commissioning an instrument that has been out of service

for a while.

5-3-7. General

Code 70

*LOAD

The load defaults code allows the instrument to be returned to the default set

up with a single operation. This can be useful when wanting to change from

one application to another.

Code 79

*CUST.D Load customer defaults. This code allows the instrument to be returned to the

factory default set, except that buffer tables (code 24,25,26) are unchanged.

5-3-8. Test and setup mode

Code 80

*TEST

Not used

Note

: attempting to change data in service code, 80 and above without the proper instructions and

equipment, can result in corruption of the instrument setup, and will impair the performance of

the unit.